<Parameter>
<ParameterFamily>hdp</ParameterFamily>
<LongName>
beam_balance_ready
</LongName>
<Name>
brt
</Name>
<AppliesTo>
<UnnumberedList>
<ListItem><TextApplies>Thermal</TextApplies></ListItem>
</UnnumberedList>
</AppliesTo>
<Description>
<DescriptionText><text>
The "beam balance ready" time specifies the time to wait after turning the
laser on and before measuring the beam with the beam balance sensor.
</text></DescriptionText>
<DescriptionText><text>
Having this value too low (ie. less than 20 seconds) will cause the stroke
profile of the beam to be measured incorrectly. Enough time is required to
allow the sensor to stabilize before any measurements are taken. The
recommended value is now 40 seconds.
</text></DescriptionText>
</Description>
<Type>
integer
</Type>
<Default>
<UnnumberedList>
<li>20</li>
</UnnumberedList>
</Default>
<Range>
<UnnumberedList>
<li>0 to 1800 [seconds]</li>
</UnnumberedList>
</Range>
<ArraySize>
</ArraySize>
<Notes>
<DescriptionText><text>
</text></DescriptionText>
</Notes>
<Partition>
cfg
</Partition>
<SeeAlso>
<UnnumberedList>
</UnnumberedList>
</SeeAlso>
<OtherLinks>
<li><Link><LinkAddress>..\..\ShowAllErrors.htm</LinkAddress><LinkTitle>Firmware 0 Error Index</LinkTitle></Link></li>
<li><Link><LinkAddress>..\..\ShowAllCommands.htm</LinkAddress><LinkTitle>Firmware 0 Command Index</LinkTitle></Link></li>
<li><Link><LinkAddress>..\..\ShowAllNVS.htm</LinkAddress><LinkTitle>Firmware 0 NVS Index</LinkTitle></Link></li>
</OtherLinks>
<Updated>
Updated: Wed Jul 16 19:36:56 2008 PST
</Updated>
<Links>
</Links>
</Parameter>